Enter the administered activity, the uptake, the effective half-life, the mean energy emitted per decay, the absorbed fraction and the organ mass to get the number of decays, the energy deposited and the absorbed dose.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MBq

Value out of range

%

Value out of range

days

Value out of range

keV

Value out of range

Value out of range

g

Value out of range

Results

Absorbed dose to the organ

66.33 Gy

A therapeutic range of dose, as in radioiodine treatments
Number of decays (× 10¹³)
4.550
Energy absorbed
1.327 J
Dose per MBq administered
0.3317 Gy/MBq
Cumulated activity
12,638 MBq·h
S value (dose per unit cumulated activity)
0.005249 Gy/(MBq·h)

The MIRD absorbed dose calculator estimates the absorbed dose to an organ from the cumulated activity, the mean energy per decay, the absorbed fraction and the organ mass, with the S value and the dose per MBq.

How the calculation works

  • Enter the activity, the uptake and the effective half-life.
  • Enter the mean energy per decay, the absorbed fraction and the mass of the organ.
  • The calculator gives the number of decays, the energy, the dose, the dose per MBq and the S value.

Example

A thyroid takes up 25 % of 200 MBq of iodine-131 with an effective half-life of 7.3 days. The number of decays is 50 MBq × 7.3 days / ln 2 = 4.550 × 10¹³. With a mean beta energy of 182 keV absorbed locally, the energy is 1.327 J. In a thyroid of 20 g, the dose is 1.327 / 0.02 = 66.33 Gy, or 0.3317 Gy/MBq administered.

Why it matters

The calculator helps nuclear medicine and medical physics students who learn internal dosimetry and the effect of activity, uptake and half-life on organ dose.

Good to know

The calculation counts only the energy deposited in the organ that holds the activity. Other organs receive a dose from gamma rays emitted there, and cross-doses require tabulated S values for each source-target pair. The model assumes uniform uptake and a single exponential clearance. For therapy, patient-specific imaging is used to refine the cumulated activity, and dose limits to critical organs are checked. For protection of staff and the public, the doses from the patient follow other rules. The figures here are for teaching and are not to be used for treatment.

The effective half-life calculator computes the residence time, the radioactive decay calculator treats decay, the radiation dose comparison calculator puts doses in context and the time, distance and shielding calculator handles exposure of staff.

Frequently asked questions

What is the MIRD method?

The Medical Internal Radiation Dose schema computes the absorbed dose to a target as the cumulated activity in the source organ times the S value: D = Ã × S. The S value contains the energy emitted per decay, the fraction of it absorbed in the target and the target mass.

What energy and absorbed fraction should I use?

For short-range particles such as beta particles, the absorbed fraction in the source organ is close to 1, and the energy is the mean particle energy per decay. For gamma rays, only a part is absorbed in the organ, and the fraction depends on the size of the organ and the photon energy.

Is this a clinical calculation?

No. Clinical dosimetry uses patient-specific imaging, several tissues, tabulated S values and software. This tool illustrates the order of magnitude and the dependence on activity, uptake, half-life and organ mass.