A calibration line turns a signal into a concentration, but the answer has an uncertainty. The calibration curve concentration calculator takes the line, the signal of the unknown and the regression statistics, and gives the concentration, its standard uncertainty, the critical t value and the confidence interval.
How the calculation works
- From your calibration regression, enter the slope m, the intercept b, the number of standards n, the residual standard deviation s y/x, the mean signal of the standards and Sxx, the sum of squared deviations of the standard concentrations.
- Enter the mean signal y0 of the unknown and the number of replicate readings k. The concentration is x0 = (y0 − b) ÷ m.
- Its standard uncertainty is s(x0) = (s y/x ÷ m) × √(1 ÷ k + 1 ÷ n + (y0 − ȳ)² ÷ (m² × Sxx)).
- The confidence interval is x0 ± t × s(x0), with t taken from the Student distribution with n − 2 degrees of freedom.
Example
A calibration line has m = 0.085, b = 0.010, n = 6 standards, s y/x = 0.006, ȳ = 0.435 and Sxx = 70. An unknown read three times averages 0.350. The concentration is (0.350 − 0.010) ÷ 0.085 = 4.0, and its standard uncertainty is 0.0506. With t = 2.776 for 4 degrees of freedom, the 95 % interval is 3.859 to 4.141.
Why it matters
Analytical chemists, environmental labs and food testers must report a concentration with its uncertainty. This is the standard inverse-prediction formula for a straight-line calibration, and it avoids the common mistake of quoting a bare number.
Good to know
The formula assumes a straight line, equal variance across the range and standards measured without error. If the line is curved or the variance grows with concentration, use a weighted or nonlinear fit. The unknown should fall inside the range of the standards.
Related tools
The linear regression calculator produces the slope, the intercept and the residual spread, and the Beer-Lambert calculator links absorbance to concentration.