Mapleson circuit fresh gas flow calculator

Choose the Mapleson circuit and the type of ventilation, and enter the weight and the minute ventilation per kilogram, to get the minute ventilation and the range of fresh gas flow that prevents rebreathing of carbon dioxide.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
kg

Value out of range

mL/kg/min

Value out of range

Results

Upper fresh gas flow

21.0 L/min

A high flow (over 15 L/min): this circuit is inefficient for this ventilation, so consider another circuit or a circle system
Lower fresh gas flow
14.0 L/min
Minute ventilation
7.0 L/min

The Mapleson circuit calculator gives the minute ventilation and the range of fresh gas flow that prevents rebreathing for the circuits A to F, in spontaneous and controlled ventilation.

How the calculation works

  • Choose the circuit and the type of ventilation.
  • Enter the weight and the minute ventilation per kilogram.
  • Read the range of fresh gas flow.

Example

For a patient of 70 kg breathing spontaneously on a Mapleson D with 100 mL/kg/min, the minute ventilation is 7.0 L/min. The fresh gas flow needed is 2 to 3 times that, or 14.0 to 21.0 L/min. A child of under 20 kg has a minimum of 3 L/min on the D, E and F circuits.

Why it matters

The calculator helps anesthesia residents, nurse anesthetists, anesthesia technicians and students who learn about breathing systems, and veterinary teams that use similar non-rebreathing circuits in small animals.

Good to know

The factors used come from classic anesthesia textbooks. Mapleson A is the most efficient in spontaneous breathing, where a flow equal to the alveolar ventilation, about 70 to 100 % of the minute ventilation, is enough, and the least efficient in controlled ventilation, where more than 2.5 times the minute ventilation or about 20 L/min is needed. The B and C circuits need about 2 to 2.5 times the minute ventilation in both modes. The D circuit, with its Bain coaxial form, and the E and F circuits need 2 to 3 times the minute ventilation in spontaneous breathing but only 1 to 2 times in controlled ventilation. High flows waste gases and anesthetic vapor, cool and dry the airway, and pollute the operating room, so the circle system is preferred when possible. The real requirement is checked with capnography: a rise of the inspired carbon dioxide above zero means rebreathing and the flow must be increased. The calculator gives an estimate and not an instruction.

The vaporizer calculator estimates the volatile agent used at a given flow, the safe apnea calculator estimates the time before desaturation, the endotracheal tube calculator sizes the tube and the minute ventilation calculator computes the ventilation.

Frequently asked questions

What are the Mapleson circuits?

They are breathing systems without a carbon dioxide absorber, classified from A to F by the position of the fresh gas inlet, the reservoir bag and the expiratory valve. They are light and simple, and used in children and for transport.

Why does the flow depend on the circuit?

Without an absorber, carbon dioxide is removed by flushing it out with fresh gas. The efficiency of the flushing depends on the layout of the circuit and on the way the patient breathes.

What is the best circuit?

The Mapleson A is the most efficient for spontaneous breathing and the D, E and F for controlled ventilation. The circle system with an absorber uses much lower flows and is the standard for adults.