Radiopharmaceutical draw volume calculator

Enter the activity and volume of the vial at calibration, the half-life, the time elapsed, the dose to inject and the delay before injection to get the current concentration, the volume to draw and the activity left in the vial.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MBq

Value out of range

mL

Value out of range

hours

Value out of range

hours

Value out of range

MBq

Value out of range

hours

Value out of range

Results

Volume to draw

6.72 mL

The vial holds enough activity for the dose
Activity of the vial now
1,166.3 MBq
Concentration now
116.63 MBq/mL
Activity to draw (decay to injection included)
783.9 MBq
Activity left in the vial after drawing
382.4 MBq

The radiopharmaceutical draw volume calculator gives the volume to draw from a vial for a prescribed activity, from the calibration, the half-life, the elapsed time and the delay before injection.

How the calculation works

  • Enter the activity and volume of the vial at calibration and the half-life.
  • Enter the time elapsed and the activity to inject with the delay.
  • Read the current concentration, the volume to draw and the activity left.

Example

A vial of technetium-99m with 1850 MBq in 10 mL at calibration is used 4 hours later. Its activity is now 1,166.3 MBq, a concentration of 116.63 MBq/mL. To inject 740 MBq in 0.5 hours, you must draw 783.9 MBq, which is 6.72 mL. About 382.4 MBq remain in the vial.

Why it matters

The calculator helps nuclear medicine technologists, radiopharmacists and students who prepare patient doses and who need to plan the use of a generator elution or a delivered vial.

Good to know

The calculation assumes a single radionuclide and a uniform concentration in the vial. Generator eluates can contain a small amount of the parent or long-lived impurities, and the activity must be checked with the dose calibrator, which is the legal reference for the dose given. Remember to allow for the activity that stays in the needle and the syringe, usually a few percent, and for the dead volume of the vial. Always label the syringe with the activity, the time and the volume, and follow local procedures for shielding, handling and waste. Pediatric doses are often calculated by weight and by minimum activity, and the same decay arithmetic applies. Time must be measured in the same unit as the half-life, here hours.

The PET SUV calculator uses the injected dose, the radioactive decay calculator handles the exponential law, the half-life calculator converts constants and the effective half-life calculator treats biological clearance.

Frequently asked questions

Why does the activity to draw exceed the dose?

If injection happens after a delay, part of the activity will decay in the syringe, so you draw a little more to deliver the prescribed activity at the time of injection: A_draw = A_inject × exp(ln 2 × delay / half-life).

Which half-lives should I use?

Technetium-99m: 6.01 hours. Fluorine-18: 1.83 hours (109.8 min). Gallium-68: 1.13 hours. Iodine-123: 13.2 hours. Indium-111: 67.3 hours. Thallium-201: 73.0 hours. Always check the data of your supplier.

Does this replace a dose calibrator?

No. The dose in each syringe must always be measured with a calibrated dose calibrator before it is given. This page helps plan the volume to draw and understand the decay.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.