Biochemists express how much enzyme they have in units, not in milligrams. The enzyme activity calculator turns a measured absorbance change per minute into enzyme activity with the extinction coefficient, the path length, the reaction and sample volumes, the dilution factor and the protein concentration. It gives the specific activity, the activity per millilitre, the activity in the cuvette, the reaction rate and the activity in katal.
How the calculation works
- Enter the change in absorbance per minute (ΔA/min) you measured on the linear part of the progress curve.
- Enter the molar extinction coefficient of the product at your wavelength, for example 6.22 mM⁻¹ cm⁻¹ for NADH at 340 nm, and the path length of the cuvette.
- Enter the total reaction volume, the volume of enzyme sample you added and its dilution factor if you diluted it.
- Enter the protein concentration of the undiluted sample to get the specific activity in units per milligram.
Example
A reaction with NADH gives ΔA/min = 0.25 in a 1 cm cuvette with ε = 6.22 mM⁻¹ cm⁻¹. The rate is 0.25 ÷ 6.22 = 0.0402 mM/min. In 1.0 mL that is 0.0402 U. The sample was diluted 10 times and 0.02 mL was added, so the activity of the undiluted sample is 0.0402 × 10 ÷ 0.02 = 20.1 U/mL. With 0.8 mg/mL protein, the specific activity is 25.1 U/mg, or 0.335 µkat/mL.
Why it matters
Activity and specific activity are the numbers used to compare preparations, follow a purification, and report the amount of enzyme in a kit or experiment. Doing the unit conversions by hand is a common source of errors in volumes and dilutions.
Good to know
The result assumes that the absorbance change comes from one product with a known extinction coefficient, that the rate is linear, and that the path length and volumes are accurate. It ignores background reactions, which should be subtracted with a blank, and inhibitors in the sample. Conditions of temperature and pH must be reported with the activity.
Related tools
The Michaelis-Menten calculator models how the rate depends on substrate, the Beer-Lambert calculator links absorbance to concentration, and the dilution calculator prepares the samples.