Skip to content
Omnikalc

Minimum detectable activity (Currie) calculator

Enter the background counts, the counting time, the detection efficiency, the emission yield and the sample size to get the critical level, the detection limit in counts and the minimum detectable activity in becquerels and per liter.

Updated on
Your inputs

Value out of range

s

Value out of range

%

Value out of range

%

Value out of range

L

Value out of range

Bq/L

Value out of range

Results

Minimum detectable activity

0.751 Bq

Critical level (counts)
46.6
Detection limit (counts)
95.7
Minimum detectable concentration
0.751 Bq/L
The method can detect the required level with margin
Minimum detectable activity in picocuries
20.3 pCi
Critical level as a net count rate
0.0777 counts/s

The minimum detectable activity calculator applies Currie’s equations to give the critical level, the detection limit and the minimum detectable activity of a radioactivity measurement, from the background counts, the counting time, the efficiency and the yield.

How the calculation works

  • Enter the background counts, the counting time and the sample size.
  • Enter the efficiency and the emission yield.
  • Read the critical level, the detection limit and the MDA.

Example

A counting system records 400 background counts in 600 s, with an efficiency of 25 % and a yield of 85 %. The critical level is 2.33 × √400 = 46.6 counts and the detection limit is 2.71 + 4.65 × 20 = 95.7 counts. The MDA is 95.7 / (600 × 0.25 × 0.85) = 0.751 Bq, or 0.751 Bq/L for a 1 L sample.

Why it matters

The calculator helps laboratory technicians, health physicists, environmental monitoring teams and students who plan low-level radioactivity measurements such as radon in water, food or soil samples.

Good to know

Currie’s equations assume that counts follow a Poisson distribution that can be approximated by a normal one, that the background is measured for the same time as the sample, and that the risks of a false positive and a false negative are both 5 percent. If the background is counted for a longer time than the sample, the formulas change and the detection limit improves. When the background is very low, for instance below ten counts, the Gaussian approximation weakens and exact Poisson methods are better. The calculation must also include decay corrections between sampling and counting, the chemical recovery and any attenuation in the sample. Report the MDA with the counting conditions so that others can reproduce it. The MDA is a property of the method, not of a sample, and it should be well below the action level.

The limit of detection calculator treats analytical chemistry blanks, the Poisson distribution calculator gives the counting statistics, the radioactive decay calculator corrects for decay and the effective half-life calculator treats body retention.

Frequently asked questions

What are the critical level and the detection limit?

The critical level is the net count above which you decide that activity is present, with a 5 % false positive risk. The detection limit is the true net count that gives a 95 % chance of detection. Currie gives Lc = 2.33√B and Ld = 2.71 + 4.65√B for a paired background of equal duration.

How do I convert counts to activity?

Divide by the counting time, the detection efficiency and the emission yield: A = N / (t × ε × y). The efficiency comes from calibration with a known source, in the same geometry as the sample.

How can I improve the minimum detectable activity?

Count longer, lower the background by shielding, increase the efficiency with a better geometry or a larger detector, or use a larger sample. The detection limit improves roughly as one over the square root of the time when the background is dominant.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.