Fluorescence quantum yield calculator

Enter the quantum yield of a reference dye, the integrated emission and the absorbance at the excitation wavelength of the sample and the reference, and the refractive indices of the solvents, to get the quantum yield of the sample.

Updated on
Your inputs

Value out of range

a.u.

Value out of range

a.u.

Value out of range

Value out of range

Value out of range

Value out of range

Value out of range

Results

Quantum yield of the sample

0.4778

Moderately fluorescent
Quantum yield as a percentage
47.8 %
Relative emission per absorbed photon
0.885 ×
Highest absorbance used
0.050
Dilute enough for the relative method
Refractive index correction
1.0409

The fluorescence quantum yield calculator gives the quantum yield of a sample by the relative method, from a reference dye, the integrated emission and absorbance of both, and the refractive indices of the solvents.

How the calculation works

  • Enter the quantum yield and the integrated emission of the reference and of the sample.
  • Enter the absorbances at the excitation wavelength and the refractive indices.
  • The calculator gives the quantum yield, its percentage, the relative emission, the highest absorbance and the index correction.

Example

A reference with a yield of 0.54 gives 1,000 units, and the sample gives 850 at the same absorbance of 0.05. The solvents have indices of 1.36 and 1.333, so the correction is 1.0409. The quantum yield is 0.54 × 0.85 × 1 × 1.0409 = 0.4778, or 47.8 %: the sample emits 0.885 times the reference per absorbed photon.

Why it matters

The calculator helps chemistry and biology students, researchers who characterize fluorescent dyes, probes or quantum dots, and laboratories that check the brightness of new materials.

Good to know

Use integrated emission spectra, the area under the whole curve, corrected for the instrument response, and not the intensity at one wavelength. Subtract the blank solvent signal. Measuring several dilutions and using the slope of the integrated emission against absorbance, rather than a single point, gives a more reliable value and checks for linearity. The refractive index correction matters when the solvents differ, for example water for the reference and ethanol or toluene for the sample. Temperature, oxygen and pH can all change the yield. Report the reference and the excitation wavelength together with the result, since the yield may depend on both.

The Beer-Lambert calculator links absorbance and concentration, the Stokes shift calculator measures the gap between absorption and emission, the photon energy calculator converts wavelength to energy and the wavelength calculator relates frequency and wavelength.

Frequently asked questions

What is the relative method?

The quantum yield is the number of photons emitted divided by the number absorbed. Instead of measuring absolute photons, the sample is compared with a reference of known yield, measured under the same conditions: Φ = Φref × (I / Iref) × (Aref / A) × (n² / nref²).

Why keep the absorbance low?

The formula assumes that the absorbed light is proportional to the absorbance, which holds below about 0.1. At higher values, the solution absorbs its own emission and the excitation light is attenuated across the cell, so the result is too low.

How to choose the reference?

Pick a standard whose absorption and emission are close to the sample, such as quinine sulfate in dilute sulfuric acid (about 0.55), fluorescein in base (0.9) or rhodamine 6G in ethanol (0.95). Use the same excitation wavelength and instrument settings for both.