Limit of detection and limit of quantification calculator for analytical methods

Enter the standard deviation of the blank and the calibration slope to get the LOD and the LOQ, and check where your measured result falls.

Updated on
Your inputs

Value out of range

Value out of range

Value out of range

Value out of range

Value out of range

Results

Limit of detection (LOD)

0.4659

Limit of quantification (LOQ)
1.4118
Measured concentration ÷ LOD
4.51
Above the limit of quantification: the result can be reported
Measured concentration ÷ LOQ
1.49

Every analytical method has a point below which it cannot see the analyte. The limit of detection calculator takes the standard deviation of the blank, the calibration slope and the ICH factors, and gives the limit of detection (LOD), the limit of quantification (LOQ) and a comparison with your measured concentration.

How the calculation works

  • Enter the standard deviation σ of the blank (or of the regression intercept) and the slope m of your calibration curve, in the same signal units.
  • The limit of detection is LOD = k × σ ÷ m, with the factor k set to 3.3 by the ICH Q2 guideline. The limit of quantification is LOQ = kq × σ ÷ m, with kq equal to 10.
  • Enter the concentration measured in your sample to see how it compares with both limits.
  • A result below the LOD is reported as not detected, one between the LOD and the LOQ as present but not quantified, and one above the LOQ as a reliable number.

Example

A calibration curve has a slope of 0.085 absorbance units per mg/L, and the blank has a standard deviation of 0.012. The LOD is 3.3 × 0.012 ÷ 0.085 = 0.4659 mg/L and the LOQ is 10 × 0.012 ÷ 0.085 = 1.4118 mg/L. A sample measured at 2.1 mg/L is 4.5 times the LOD and 1.5 times the LOQ, so it can be reported.

Why it matters

Method validation in analytical chemistry, pharmaceutical quality control, environmental testing and clinical labs all require the LOD and LOQ. Reporting results below these limits without care can mislead regulators and customers.

Good to know

This signal-to-slope approach is one of several accepted ways to estimate the limits, and others use signal-to-noise ratios or visual evaluation. The values should be confirmed by measuring samples near the computed limits. The units of σ and the slope must be consistent.

The standard error calculator gives the spread of an average, and the coefficient of variation calculator measures precision in relative terms.

Frequently asked questions

What is the difference between the LOD and the LOQ?

The limit of detection is the smallest concentration that can be told apart from a blank with reasonable confidence. The limit of quantification is the smallest concentration that can be measured with acceptable precision and accuracy. The LOQ is always higher than the LOD, usually about three times higher.

Where do the factors 3.3 and 10 come from?

The ICH Q2 guideline sets the detection limit at 3.3 times the standard deviation of the response divided by the slope, and the quantification limit at 10 times. The 3.3 comes from about 1.645 for each of two error types at roughly 5 percent each, and the 10 gives a relative error near 10 percent.

Which standard deviation should I use?

The usual choices are the standard deviation of replicate blanks, the residual standard deviation of the regression line, or the standard deviation of the intercept of a low-level calibration. The guideline allows any of them as long as you state it and the method is validated.

Can I report a result below the LOQ?

Often yes, but as an estimate or as a trace, not as a precise value. Regulations differ, so follow the reporting rule of your method, laboratory or accreditation body.