Leak rate calculator: measure a vacuum leak from the pressure rise

Enter the volume of a vacuum chamber and the pressures at the start and the end of a rate-of-rise test to get the rate of pressure rise and the leak rate in torr·L/s, mbar·L/s, sccm and liters of gas per day.

Updated on
Your inputs
L

Value out of range

torr

Value out of range

torr

Value out of range

s

Value out of range

Results

Leak rate Q = V ΔP / Δt

0.00267 torr·L/s

Rate of pressure rise
8.0 mtorr/min
Leak rate
0.00356 mbar·L/s
Leak rate
0.211 sccm
Gas entering per day, at atmospheric pressure
0.303 L/day

A rate-of-rise test measures a vacuum leak from the pressure that builds up in a closed chamber. The leak rate calculator gives the rate of pressure rise, the leak rate in torr·L/s, mbar·L/s, sccm and the volume of gas that enters each day.

How the calculation works

  • Close the valve between the chamber and the vacuum pump, then record the pressure at the start, the pressure at the end and the time that passed.
  • Enter the volume of the chamber and the pipes that are cut off from the pump, in liters.
  • The calculator computes the rate of pressure rise ΔP / Δt and the leak rate Q = V × ΔP / Δt, in torr·L/s.
  • It converts the leak rate to mbar·L/s and to standard cubic centimeters per minute (sccm), and shows the volume of gas at atmospheric pressure that enters in a day.

Example

A chamber of 20 L is pumped to 0.100 torr, the pump valve is closed, and the pressure rises to 0.180 torr in 600 s. The rate of rise is 80 mtorr in 10 min = 8.0 mtorr/min. The leak rate is 20 × 0.08 / 600 = 0.00267 torr·L/s, which is 0.00356 mbar·L/s or 0.211 sccm. That is about 0.303 L of air per day at atmospheric pressure. If the leak was a virtual leak or outgassing, the rate would fall with time.

Why it matters

The calculator helps vacuum technicians, scientists, semiconductor and coating equipment engineers, and makers of vacuum furnaces and chambers to check the tightness of a system, to decide if a leak is acceptable, and to find if a long pump-down is caused by a leak. It gives a standard figure to compare with the specification.

Good to know

Wait for the pressure to stabilize after closing the valve, and measure over a long enough time. Repeat the test to see if the rate changes. Include the volume of the pipes up to the closed valve. Use a helium leak detector to find the location of the leak, not only its size.

The vacuum pump-down time calculator gives the time to reach a pressure, the ideal gas calculator relates pressure, volume and moles, the tank volume calculator gives the volume of a vessel and the pressure tank calculator covers pressure vessels.

Frequently asked questions

What is a rate-of-rise test?

In a rate-of-rise test, the vacuum chamber is pumped to a low pressure, the valve to the pump is closed, and the pressure is recorded over time. The gas that enters, through a real leak or from the walls, raises the pressure. The leak rate is the volume of the chamber times the slope of the pressure curve: Q = V dP/dt. It is the simplest way to measure the tightness of a system, without a leak detector.

How can I tell a real leak from outgassing?

A real leak makes a straight line: the pressure rises at a constant rate, as long as it is far below atmospheric pressure. Outgassing from the walls, water vapor above all, gives a curve that bends and flattens with time, since the surface desorbs less and less. A virtual leak, from trapped gas in a blind hole or under a bolt, shows a rate that falls more slowly. Compare several tests and a baking of the chamber.

What are the units of the leak rate?

The leak rate is a throughput, a pressure times a volume per unit of time: torr·L/s, mbar·L/s or Pa·m³/s. One torr·L/s is 1.333 mbar·L/s, or 0.1333 Pa·m³/s. The sccm, standard cubic centimeters per minute, is the equivalent volume of gas at standard conditions: 1 torr·L/s is about 79 sccm. A leak detector with helium gives a rate for helium, which is about 2.7 times that of air by flow in the molecular regime.

What are the limits of this calculator?

The calculation assumes a constant rate over the test and a uniform pressure in the volume. It includes all the sources of gas: leaks, outgassing and permeation. For a system with a leak, the result is the sum of them. The accuracy depends on the pressure gauge, which must be accurate in the range of the test. Use a longer test or a larger volume for a small leak.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.