Doppler ultrasound velocity calculator

Enter the transmitted frequency, the measured Doppler shift, the angle between the beam and the flow and the pulse repetition frequency to get the blood velocity, the aliasing limit and the error caused by a wrong angle.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MHz

Value out of range

kHz

Value out of range

°

Value out of range

kHz

Value out of range

m/s

Value out of range

Results

Blood velocity

0.801 m/s

Blood velocity in centimeters per second
80 cm/s
Maximum velocity before aliasing
0.924 m/s
Share of the aliasing limit used
87 %
Close to the aliasing limit
Velocity error for a 5° angle error
18 %
Acceptable sensitivity to the angle

The Doppler ultrasound velocity calculator converts a Doppler shift into a blood velocity using the transmitted frequency, the beam angle and the speed of sound, and checks the aliasing limit of the pulse repetition frequency.

How the calculation works

  • Enter the frequency, the Doppler shift and the angle.
  • Enter the PRF and the speed of sound.
  • Read the velocity, the aliasing limit and the angle error.

Example

A Doppler shift of 2.6 kHz at 5 MHz with a beam angle of 60° gives v = 1540 × 2600 / (2 × 5 000 000 × 0.5) = 0.801 m/s, or 80 cm/s. With a PRF of 6 kHz the aliasing limit is 0.924 m/s, so the measurement uses 87 % of it. A 5° error would change the velocity by 18 %.

Why it matters

The calculator helps sonographers, cardiologists, vascular technologists and students who learn how Doppler works, and engineers who design ultrasound systems.

Good to know

The equation assumes a constant speed of sound of 1540 m/s, a flow parallel to the vessel wall and a known angle between the flow and the beam. Real flow is not uniform, so the spectrum contains a range of velocities, and the maximum envelope is normally used. Continuous wave Doppler has no aliasing limit but gives no depth information, while pulsed Doppler locates the signal at the cost of a velocity ceiling. Wall filters remove low-frequency tissue motion and may hide slow flow. Never measure velocities with an angle near 90°, because the cosine is close to zero and the result is unreliable. Use the angle correction cursor aligned with the flow and report the angle with the measurement.

The resistive and pulsatility index calculator describes the shape of the waveform, the pulse-echo timing calculator links depth and PRF, the mechanical index calculator checks the acoustic output and the Doppler effect calculator treats sound sources in air.

Frequently asked questions

What is the Doppler equation?

The shift in frequency of the echoes from moving blood is Δf = 2 f0 v cos θ / c, so the velocity is v = c Δf / (2 f0 cos θ). The factor of 2 appears because the shift happens twice, once on the way in and once on the way back.

What is aliasing?

Pulsed Doppler samples the signal at the pulse repetition frequency. A shift above half the PRF (the Nyquist limit) is wrongly displayed as a lower or negative velocity, which gives the typical wrap-around of the spectrum.

Why is the angle so important?

The cosine falls quickly beyond 60°, so a small angle error produces a large velocity error. Operators therefore keep the angle below 60° and always correct for it, and angles above 70° should not be used for quantitative velocities.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.