Ultrasound mechanical index (MI) calculator

Enter the peak rarefactional pressure measured in water, the center frequency, the depth of the focus and the attenuation coefficient to get the derated pressure, the mechanical index and its comparison with the usual regulatory limits.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MPa

Value out of range

MHz

Value out of range

cm

Value out of range

dB/cm/MHz

Value out of range

Results

Mechanical index

0.52

Below the general limit, above the 0.23 limit for ophthalmic use
Derated peak rarefactional pressure
0.968 MPa
Share of the general limit of 1.9
27 %
Derating factor
0.484
Share of the ophthalmic limit of 0.23
225 %

The ultrasound mechanical index calculator gives the derated pressure and the mechanical index from the peak rarefactional pressure, the frequency, the focus depth and the attenuation coefficient.

How the calculation works

  • Enter the pressure measured in water, the frequency and the focus depth.
  • Enter the attenuation coefficient.
  • Read the derated pressure, the MI and the share of the limits.

Example

A pressure of 2.0 MPa in water at 3.5 MHz is derated for 6 cm of tissue at 0.3 dB/cm/MHz: the factor is 10^(−0.3 × 3.5 × 6 / 20) = 0.484, giving 0.968 MPa. The mechanical index is 0.968 / √3.5 = 0.52, which is 27 % of the general limit of 1.9 and 225 % of the ophthalmic limit of 0.23.

Why it matters

The calculator helps sonographers, biomedical engineers, medical physicists and students who study the safety of diagnostic ultrasound, and teams who use contrast agents or scan the eye or a fetus.

Good to know

The mechanical index is not a measure of the dose, but an indicator of the likelihood of cavitation. It rises with the pressure and falls with the frequency, so low frequency probes at high output carry the largest values. Gas microbubble contrast agents lower the threshold, which is why contrast exams use a low MI, often below 0.3, to avoid destroying the bubbles. The thermal index is a separate indicator of heating. Regulators allow up to 1.9 for most applications and 0.23 for ophthalmic use. Follow the ALARA principle by using the lowest output and the shortest exposure that gives the diagnostic information, especially for fetal exams. The displayed MI must be read on the screen of the scanner.

The pulse-echo timing calculator links depth and PRF, the resolution calculator treats the beam, the Doppler velocity calculator gives blood speed and the acoustic impedance calculator treats reflection at boundaries.

Frequently asked questions

What is the mechanical index?

The mechanical index is the derated peak negative pressure in megapascals divided by the square root of the center frequency in megahertz: MI = pr,0.3 / √f. It estimates the risk of non-thermal effects such as cavitation, where gas bubbles form or collapse in the tissue.

Why derate the pressure?

The pressure is measured in water, which does not absorb the sound. In tissue the beam is attenuated, usually taken as 0.3 dB/cm/MHz, so the pressure at depth is lower. Standards require this derating for the displayed value.

Is the displayed MI the same as this one?

The scanner displays an MI computed from its own calibration data. This page is for learning and for checking the order of magnitude, and it does not replace the display or the manufacturer’s data.