The ultrasound pulse-echo timing calculator gives the round-trip time, the maximum PRF, the frame rate and the unambiguous depth of an ultrasound image from the depth, the speed of sound and the scan lines.
How the calculation works
- Enter the depth and the speed of sound.
- Enter the number of lines and focal zones and a PRF.
- Read the round-trip time, the frame rate and the unambiguous depth.
Example
An image of 12 cm at 1540 m/s: the echo returns after 2 × 0.12 / 1540 = 155.8 µs, so the maximum PRF is 6,417 Hz. With 128 lines and 1 focal zone, the frame rate is 50.1 frames per second. A PRF of 4000 Hz allows an unambiguous depth of 1540 / (2 × 4000) = 19.2 cm.
Why it matters
The calculator helps sonography students, technologists, biomedical engineers and developers of ultrasound systems who need to understand the trade-off between depth, line density and speed.
Good to know
The calculation assumes a constant speed of sound of 1540 m/s, which is the average value used by scanners to convert time to depth. Fat is slower, around 1450 m/s, and bone is faster, so structures can appear displaced. Real systems use parallel receive beamforming, plane waves or coded excitation to increase the frame rate without changing the physics of the round trip. Time gain compensation and the duty factor are also tuned in practice. For Doppler, a higher PRF increases the velocity range but lowers the unambiguous depth. When scanning deep, a lower frame rate is accepted for a better image. Use this page to understand why the settings of the scanner interact.
Related tools
The resolution calculator describes the beam, the mechanical index calculator checks safety, the Doppler velocity calculator uses the PRF and the acoustic impedance calculator treats reflections.