The entrance skin dose calculator gives the incident air kerma, the entrance skin dose and the dose area product of a radiograph from the tube output, the mAs, the distance and the backscatter factor.
How the calculation works
- Enter the tube output, the mAs and the focus to skin distance.
- Enter the backscatter factor, the field size and the reference level.
- Read the ESD, the DAP and the share of the reference level.
Example
A tube with an output of 0.05 mGy/mAs at 1 m used at 20 mAs and a distance of 90 cm gives an incident air kerma of 0.05 × 20 × (100 / 90)² = 1.23 mGy. With a backscatter factor of 1.35 the skin dose is 1.67 mGy. A field of 30 × 40 cm gives a DAP of 1.48 Gy·cm², and the dose is 17 % of a reference level of 10 mGy.
Why it matters
The calculator helps radiographers, medical physicists and quality control teams who check patient doses against diagnostic reference levels, and students who learn the dose quantities of radiography.
Good to know
The estimate assumes that the tube output is known for the kVp and the filtration used, and that the field is fully on the patient. The output varies with the generator, the age of the tube and the added filtration, so measure it regularly. Patient dose depends on thickness: a large patient needs a higher mAs and gets a higher skin dose, but the dose at the receptor stays constant. The DAP is easier to measure in practice with a meter on the collimator and is the basis of most dose audits. The effective dose factor given here is only a rough guide that varies by examination, so use published conversion factors for the body part examined. Follow the ALARA principle and compare your results with local or national reference levels.
Related tools
The exposure adjustment calculator changes the technique, the CT dose calculator treats scanners, the inverse square law calculator shows the effect of distance and the dose comparison calculator puts a dose in context.