The radiographic exposure adjustment calculator gives the new mAs that keeps the same receptor exposure when you change the kVp, the distance or the grid, with the 15 percent rule, the inverse square law and the grid factor.
How the calculation works
- Enter the mAs, kVp, distance and grid of the original technique.
- Enter the new kVp, distance and grid.
- Read the new mAs and the factor from each change.
Example
A technique of 20 mAs at 80 kVp, 100 cm and an 8:1 grid is changed to 92 kVp (15 % more) and 150 cm with the same grid. The kVp factor is 0.500, the distance factor is 2.250, the grid factor is 1.000, so the new technique is 22.50 mAs, an overall factor of 1.125.
Why it matters
The calculator helps radiography students and technologists who must adapt a technique to a different room, a mobile unit, a patient too large for the usual technique or a change of grid.
Good to know
These are rules of thumb, not exact physics. The 15 percent rule changes the contrast and the patient dose, the inverse square law assumes a point source and no air attenuation, and the grid factor depends on the grid, the kVp and the field size. Digital systems tolerate a wider range of exposures, but exposure index values still show whether the receptor received enough signal. Never use a higher mAs just to compensate for a bad technique without thinking about dose, and follow the ALARA principle by using the lowest exposure that gives a diagnostic image. When the original technique was not well adapted, the adjusted one will not be either, so start from a good reference technique.