Gamma index calculator (dose difference and DTA)

Enter the reference and measured dose, the distance between the points, the dose difference and distance-to-agreement criteria and the number of points to get the gamma value of the pair, whether it passes, and the gamma pass rate against the action levels.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
Gy

Value out of range

Gy

Value out of range

mm

Value out of range

%

Value out of range

mm

Value out of range

Gy

Value out of range

Value out of range

Value out of range

%

Value out of range

Results

Gamma value of the pair of points

0.972

The point passes: gamma is 1 or lower
Dose difference
2.50 %
Gamma pass rate
96.5 %
At or above 95 %: the plan is delivered as planned
Dose component of gamma
0.833
Distance component of gamma
0.500

The gamma index calculator gives the gamma value of a pair of points from the dose difference, the distance and the criteria, and compares a pass rate with the tolerance and action levels of patient-specific quality assurance.

How the calculation works

  • Enter the reference and measured doses and the distance.
  • Enter the criteria and the normalization.
  • Enter the points evaluated and passing and read the gamma and the pass rate.

Example

A reference dose of 2.00 Gy against a measured 2.05 Gy at a distance of 1.5 mm, with a global normalization to 2 Gy, gives a dose difference of 2.50 %. With criteria of 3 % and 3 mm the gamma value is √[(1.5 / 3)² + (2.50 / 3)²] = 0.972, so the point passes. A pass rate of 965 points out of 1000 is 96.5 %.

Why it matters

The calculator helps medical physicists, dosimetrists and students who run patient-specific quality assurance of IMRT and VMAT plans and who want to understand what a gamma analysis computes.

Good to know

A gamma analysis depends on the choice of criteria, the normalization, the low-dose threshold, the dose grid resolution and the type of detector, and the same plan can pass with one choice and fail with another. A high pass rate does not guarantee that the plan is good, and a low pass rate does not always mean a delivery error, since a detector resolution problem or a setup error can be the cause. Analysis in three dimensions and the clinical impact on the dose-volume histograms are more informative than the pass rate alone. The AAPM TG-218 recommendations describe the tolerance and action limits and the way to investigate. This calculation covers one pair of points, which is the building block of the full analysis.

The conformity index calculator evaluates a plan, the monitor unit calculator links dose and machine output, the ion chamber correction calculator treats absolute dose measurement and the standard deviation calculator treats the spread of measurements.

Frequently asked questions

What is the gamma index?

The gamma index combines a dose difference criterion and a distance-to-agreement criterion in a single number: γ = √[(Δd / DTA)² + (ΔD / DD)²]. A point passes when γ is 1 or less, that is, when it falls inside an ellipsoid defined by the two criteria.

Why does the real analysis search around the point?

For each measured point the software looks for the reference point that gives the smallest γ. This page computes γ for one chosen pair, which is the simple case, and shows how each term contributes.

Which criteria and levels are used?

Common criteria are 3 % and 3 mm, or 3 % and 2 mm, with a global normalization and a threshold of 10 % of the maximum dose. The AAPM TG-218 report gives a tolerance of 95 % and an action level of 90 % for 3 %/2 mm patient-specific QA.