The gamma index calculator gives the gamma value of a pair of points from the dose difference, the distance and the criteria, and compares a pass rate with the tolerance and action levels of patient-specific quality assurance.
How the calculation works
- Enter the reference and measured doses and the distance.
- Enter the criteria and the normalization.
- Enter the points evaluated and passing and read the gamma and the pass rate.
Example
A reference dose of 2.00 Gy against a measured 2.05 Gy at a distance of 1.5 mm, with a global normalization to 2 Gy, gives a dose difference of 2.50 %. With criteria of 3 % and 3 mm the gamma value is √[(1.5 / 3)² + (2.50 / 3)²] = 0.972, so the point passes. A pass rate of 965 points out of 1000 is 96.5 %.
Why it matters
The calculator helps medical physicists, dosimetrists and students who run patient-specific quality assurance of IMRT and VMAT plans and who want to understand what a gamma analysis computes.
Good to know
A gamma analysis depends on the choice of criteria, the normalization, the low-dose threshold, the dose grid resolution and the type of detector, and the same plan can pass with one choice and fail with another. A high pass rate does not guarantee that the plan is good, and a low pass rate does not always mean a delivery error, since a detector resolution problem or a setup error can be the cause. Analysis in three dimensions and the clinical impact on the dose-volume histograms are more informative than the pass rate alone. The AAPM TG-218 recommendations describe the tolerance and action limits and the way to investigate. This calculation covers one pair of points, which is the building block of the full analysis.
Related tools
The conformity index calculator evaluates a plan, the monitor unit calculator links dose and machine output, the ion chamber correction calculator treats absolute dose measurement and the standard deviation calculator treats the spread of measurements.