Radiotherapy conformity and homogeneity index calculator

Enter the target volume, the volume covered by the prescription isodose, the total prescription isodose volume, the maximum dose, the prescription dose and the volume of the half-prescription isodose to get the RTOG and Paddick conformity indices, the homogeneity index, the coverage and the gradient index.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
cc

Value out of range

cc

Value out of range

cc

Value out of range

Gy

Value out of range

Gy

Value out of range

cc

Value out of range

Results

RTOG conformity index

1.24

Between 1 and 2: the plan complies with the RTOG criterion
Paddick conformity index
0.74
An acceptable conformity
Coverage of the target
96.0 %
Homogeneity index (maximum dose / prescription dose)
1.25
Gradient index (half-prescription volume / prescription volume)
3.06
Selectivity (target covered / prescription isodose volume)
77.4 %

The conformity and homogeneity index calculator gives the RTOG conformity index, the Paddick index, the homogeneity index, the coverage and the gradient index of a radiotherapy plan from the volumes and doses.

How the calculation works

  • Enter the target volume and the volume covered.
  • Enter the prescription isodose volume, the doses and the half-prescription volume.
  • Read the indices and the coverage.

Example

A target of 25 cc has 24 cc covered by the prescription isodose, which encloses 31 cc in total. The RTOG index is 31 / 25 = 1.24. The Paddick index is 24² / (25 × 31) = 0.74. The coverage is 96.0 %. With a prescription of 24 Gy and a maximum of 30 Gy the homogeneity index is 1.25. A half-prescription volume of 95 cc gives a gradient index of 3.06.

Why it matters

The calculator helps medical physicists, dosimetrists, radiation oncologists and students who evaluate treatment plans, especially for stereotactic radiosurgery and stereotactic body radiotherapy.

Good to know

The indices summarize the plan in one number each and do not replace the review of the dose distribution and the dose-volume histograms. A high conformity index may be obtained by sacrificing the coverage, so always check the coverage at the same time. Definitions vary between centers and trials, and the Paddick and RTOG indices are used together or with the gradient index. For small targets the indices depend heavily on the calculation grid resolution and on the way the volumes are contoured. Organs at risk must be evaluated separately with their own constraints. Use the same definition when comparing plans and say which one is used when you report the result.

The proton range calculator treats particle beams, the gamma index calculator compares dose distributions, the monitor unit calculator links dose and machine output and the TCP calculator links dose to control.

Frequently asked questions

What do the conformity indices measure?

They compare the volume enclosed by the prescription isodose with the target. The RTOG index is the isodose volume divided by the target volume and ideally is between 1 and 2. The Paddick index also accounts for the overlap, CI = TVPI² / (TV × VPI), and is 1 for a perfect plan.

What is the homogeneity index?

The RTOG homogeneity index is the maximum dose divided by the prescription dose. A value up to 2 is acceptable in radiosurgery and up to 1.1 to 1.2 is typical in conventional fractionated plans. It shows how uniform the target dose is.

What is the gradient index?

It is the ratio of the volume of the half-prescription isodose to the volume of the prescription isodose. It measures how steeply the dose falls outside the target; lower is better, typically 3 to 4 for single-fraction radiosurgery of small targets.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.