Monitor unit calculator (SSD technique)

Enter the prescribed dose at depth, the percentage depth dose, the output factor, the calibration of the machine, the treatment distance and any tray or wedge factor to get the monitor units to set on the linear accelerator.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
cGy

Value out of range

%

Value out of range

Value out of range

cGy/MU

Value out of range

cm

Value out of range

cm

Value out of range

cm

Value out of range

Value out of range

MU/min

Value out of range

Results

Monitor units

299.4

Dose at the depth of maximum
299.4 cGy
Dose at dmax in a usual range for one fraction
Beam-on time
0.50 min
Inverse-square factor
1.0000
Monitor units without tray or wedge
299.4

The monitor unit calculator converts a prescribed dose at depth into the monitor units to set on the machine, for a fixed source-to-surface distance technique, with the PDD, the output factor and tray or wedge factors.

How the calculation works

  • Enter the prescribed dose and the percentage depth dose.
  • Enter the output factor, the calibration and the distances.
  • The calculator gives the monitor units, the dose at dmax, the beam-on time and the inverse-square factor.

Example

A prescription of 200 cGy at depth with a PDD of 66.8 %, an output factor of 1.00, a calibration of 1.000 cGy/MU and the reference distance of 100 cm needs 200 / 0.668 = 299.4 MU. The dose at the depth of maximum is 299.4 cGy, and at 600 MU/min the beam is on for 0.50 min.

Why it matters

The calculator helps radiation therapy students, dosimetry trainees and medical physicists who teach the basic factors of a dose calculation.

Good to know

Real calculations also include the phantom scatter and collimator scatter factors, the effect of the field shape, off-axis ratios, and heterogeneity corrections, and modern treatments use computer planning, not manual calculations. Always keep the units straight, because the calibration in cGy per monitor unit and the dose in cGy must match for the result to be meaningful. Isocentric techniques use tissue-maximum ratios in place of percentage depth doses. The PDD depends on the energy, the field size, the depth and the distance. Make sure that all the factors come from the same beam data set. Every manual check must be independent and reviewed. This page is meant for education only.

The BED and EQD2 calculator compares fractionation schedules, the Mayneord F factor calculator corrects PDD for distance, the inverse square law calculator shows the distance effect and the equivalent fractionation calculator changes dose per fraction.

Frequently asked questions

How are monitor units computed?

In the fixed source-to-surface distance technique, MU = prescribed dose / (calibration × PDD / 100 × output factor × tray factor × inverse-square factor). The calibration is the dose per monitor unit at the depth of maximum for a reference field.

What is the inverse-square factor for?

If the treatment distance differs from the reference, the dose at depth of maximum changes with the square of the distance from the source, here measured to the depth of maximum dose. This calculator uses that correction.

Is this a clinical calculator?

No. A real calculation uses the validated data of the specific machine and beam and is checked by a medical physicist and a second independent calculation. Use this page for teaching and to understand the factors.