The ion chamber corrections calculator gives the PTP, Ppol and Pion correction factors and the absorbed dose to water from the raw readings, the temperature and pressure, the calibration coefficient and the beam quality factor.
How the calculation works
- Enter the raw readings at both polarities and at two voltages.
- Enter the temperature, the pressure and the coefficients.
- Read the corrections and the dose to water.
Example
At 21 °C and 100.2 kPa the temperature-pressure factor is 294.2 / 295.2 × 101.33 / 100.2 = 1.0079. Readings of 20.00 nC and 19.96 nC at the two polarities give a polarity factor of 0.9990. In a pulsed beam, 20.00 and 19.84 nC at 300 V and 150 V give a recombination factor of 1.0081. With 0.054 Gy/nC and a quality factor of 0.993, the dose is 1.0886 Gy.
Why it matters
The calculator helps medical physics students, physicists and technologists who learn the corrections of absolute dosimetry with ion chambers and who want to see the size of each correction.
Good to know
The corrections are usually small, within one or two percent, but they add up, and they must be applied consistently with the protocol used. The pressure must be the absolute pressure at the site, not a value reduced to sea level, which is a common mistake. The chamber must be in thermal equilibrium with the room, which takes some minutes after it is moved. The polarity correction depends on the sign of the reading, and the recombination formulas assume the measurements in the linear region of the saturation curve. The calibration coefficient comes from a calibration laboratory with its own uncertainty. Use this page to understand the structure of the calculation and not for patient dosimetry.