Ion chamber corrections calculator (PTP, Ppol, Pion)

Enter the raw readings at two polarities and two voltages, the temperature and pressure, the calibration coefficient and the beam quality factor to get the correction factors PTP, Ppol and Pion and the absorbed dose to water.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
nC

Value out of range

nC

Value out of range

nC

Value out of range

V

Value out of range

V

Value out of range

°C

Value out of range

kPa

Value out of range

Gy/nC

Value out of range

Value out of range

Results

Absorbed dose to water

1.0886 Gy

Temperature-pressure factor PTP
1.0079
Polarity factor Ppol
0.9990
Recombination factor Pion
1.0081
A usual recombination correction
Product of the three corrections
1.0150
Absorbed dose to water in centigrays
108.86 cGy

The ion chamber corrections calculator gives the PTP, Ppol and Pion correction factors and the absorbed dose to water from the raw readings, the temperature and pressure, the calibration coefficient and the beam quality factor.

How the calculation works

  • Enter the raw readings at both polarities and at two voltages.
  • Enter the temperature, the pressure and the coefficients.
  • Read the corrections and the dose to water.

Example

At 21 °C and 100.2 kPa the temperature-pressure factor is 294.2 / 295.2 × 101.33 / 100.2 = 1.0079. Readings of 20.00 nC and 19.96 nC at the two polarities give a polarity factor of 0.9990. In a pulsed beam, 20.00 and 19.84 nC at 300 V and 150 V give a recombination factor of 1.0081. With 0.054 Gy/nC and a quality factor of 0.993, the dose is 1.0886 Gy.

Why it matters

The calculator helps medical physics students, physicists and technologists who learn the corrections of absolute dosimetry with ion chambers and who want to see the size of each correction.

Good to know

The corrections are usually small, within one or two percent, but they add up, and they must be applied consistently with the protocol used. The pressure must be the absolute pressure at the site, not a value reduced to sea level, which is a common mistake. The chamber must be in thermal equilibrium with the room, which takes some minutes after it is moved. The polarity correction depends on the sign of the reading, and the recombination formulas assume the measurements in the linear region of the saturation curve. The calibration coefficient comes from a calibration laboratory with its own uncertainty. Use this page to understand the structure of the calculation and not for patient dosimetry.

The monitor unit calculator links machine output and dose, the gamma index calculator compares dose distributions, the Mayneord factor calculator corrects the depth dose for distance and the BED calculator compares fractionation schedules.

Frequently asked questions

What is the temperature-pressure correction?

An unsealed chamber contains air whose mass changes with the temperature and the pressure. The reading is corrected to the reference conditions of the calibration (22 °C and 101.33 kPa): PTP = (273.2 + T) / 295.2 × 101.33 / P.

What are the polarity and recombination corrections?

The polarity correction compares readings at both polarities, Ppol = (|M+| + |M−|) / (2 |M|). The recombination correction accounts for ions that recombine before they are collected, from readings at two voltages, with different formulas for pulsed and continuous beams.

Is this a complete calibration?

No. It follows the logic of protocols such as AAPM TG-51 or IAEA TRS-398, but a real calibration also requires the beam quality determination, the positioning, the electrometer correction and the traceability of the chamber. It is performed by a medical physicist.