The contrast to noise ratio calculator gives the CNR, the SNR, the relative contrast and the dose factor needed to reach a target CNR, from the mean signals of a lesion and of the background and the noise.
How the calculation works
- Enter the signal in the lesion and in the background.
- Enter the noise and the target CNR.
- Read the CNR, the SNR and the dose factor.
Example
A lesion with a mean signal of 120 on a background of 100 and a noise of 8 has a CNR of 20 / 8 = 2.50 and an SNR of 15.00. The relative contrast is 20.0 %. To reach a CNR of 5, the dose must be multiplied by (5 / 2.5)² = 4.00, or the noise reduced to 4.00.
Why it matters
The calculator helps medical physicists, radiographers and students who compare imaging protocols in CT, radiography, ultrasound, MRI or nuclear medicine, and anyone who measures regions of interest on an image.
Good to know
Measure the regions of interest on a uniform part of the image and avoid edges, vessels and artifacts. The noise should be the standard deviation of a uniform region, not of the lesion itself. In CT the values depend on the reconstruction kernel and on iterative algorithms, which change the noise texture without changing the dose. In MRI and ultrasound the noise is not strictly quantum noise, so the square law for the dose does not apply, but the CNR is still useful to compare settings. A CNR above the Rose criterion is a guide, not a guarantee, since the visibility of a lesion also depends on its size, shape, and the viewing conditions. Use the same measurement method when comparing two protocols.
Related tools
The CT dose calculator gives the scanner output, the signal to noise ratio calculator treats electronic signals, the exposure adjustment calculator changes the radiographic technique and the standard deviation calculator computes the noise of a region.