Modulation transfer function (MTF) calculator

Enter the full width at half maximum of the line spread function, a spatial frequency and the pixel pitch to get the MTF at that frequency, the frequencies at 50 % and 10 % modulation, and the MTF at the Nyquist frequency.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
mm

Value out of range

lp/mm

Value out of range

mm

Value out of range

Results

MTF at the chosen frequency

41.1 %

Reduced contrast at this frequency
Frequency at 50 % MTF
2.21 lp/mm
Frequency at 10 % MTF
4.02 lp/mm
Nyquist frequency of the pixel pitch
3.33 lp/mm
MTF at the Nyquist frequency
20.6 %

The MTF calculator gives the modulation transfer function of a Gaussian imaging system from the FWHM of its line spread function, the frequencies at 50 % and 10 % modulation, and the MTF at the Nyquist frequency of the pixel pitch.

How the calculation works

  • Enter the FWHM of the line spread function.
  • Enter a spatial frequency and the pixel pitch.
  • Read the MTF, the 50 % and 10 % frequencies and the Nyquist values.

Example

A system with a line spread function of 0.2 mm FWHM has σ = 0.2 / 2.355 = 0.0849 mm. At 2.5 lp/mm its MTF is 41.1 %. The MTF is 50 % at 2.21 lp/mm and 10 % at 4.02 lp/mm. A pixel pitch of 0.15 mm gives a Nyquist frequency of 3.33 lp/mm, where the MTF is 20.6 %.

Why it matters

The calculator helps medical physicists, radiology technologists, engineers who design detectors and students who learn image quality, as well as photographers and microscopists who compare lenses and sensors.

Good to know

Real MTF curves are measured with an edge, a slit or a bar pattern, and they are rarely Gaussian, especially for screens and scintillators with a long tail. The Gaussian model is a convenient approximation for comparing the effect of blur sources: for independent blurs the widths add in quadrature, so a focal spot, a detector and motion blur can be combined by taking the square root of the sum of squares. The 10 % frequency is often taken as the limiting resolution, while the 50 % frequency better reflects the practical sharpness. Sampling and display also affect the final image, so a high MTF alone does not guarantee a sharp picture. Use the same units for the width and the frequency, millimeters and line pairs per millimeter.

The geometric unsharpness calculator gives the blur of the focal spot, the contrast to noise calculator describes noise, the pixel density calculator treats displays and the normal distribution calculator shows the Gaussian shape.

Frequently asked questions

What is the MTF?

The modulation transfer function describes how much of the contrast of a pattern the imaging system keeps at each spatial frequency, from 100 % for coarse detail down to zero for the finest detail. It is the standard measure of the sharpness of an imaging system.

Why a Gaussian model?

A Gaussian line spread function is a good first model of blur from several independent sources. Its Fourier transform is also a Gaussian, MTF(f) = exp(−2π²σ²f²) with σ = FWHM / 2.355. Real systems have tails and need measured curves.

What is the Nyquist frequency?

It is half the sampling frequency, 1 / (2 × pixel pitch). Detail finer than this cannot be recorded correctly and appears as aliasing. A good detector has a low MTF at the Nyquist frequency to limit aliasing.