The nuclear medicine image counts calculator gives the counts per pixel, the total counts, the count rate and the acquisition time needed for a target image noise, or the noise reached in a chosen time, from the activity and the system sensitivity.
How the calculation works
- Enter the target noise and the pixels of the region.
- Enter the activity and the sensitivity.
- Read the time and the noise in a chosen time.
Example
A noise of 5 % per pixel needs (100 / 5)² = 400 counts per pixel, so 800,000 counts over 2000 pixels. With 50 MBq in the region and 70 cps/MBq, the count rate is 3,500 cps, and the acquisition takes 229 s, or 3.8 min. In 300 s the noise would be 4.36 %.
Why it matters
The calculator helps nuclear medicine technologists, physicists and students who plan acquisitions and want to understand the link between activity, time and image noise.
Good to know
The model uses Poisson noise only, with no correction for background, scatter, attenuation or the image processing, which includes smoothing that reduces the visible noise at the expense of resolution. The sensitivity depends on the collimator, the photon energy, the depth of the source and the energy window. Only the counts in the region count, so a bigger region needs more counts for the same noise per pixel but the same noise for the total counts. The motion of the patient increases with the scan time, and decay during a long acquisition reduces the count rate, especially for short-lived radionuclides, so allow a margin. The time is only a planning estimate and does not replace the local protocol.
Related tools
The PET count rate calculator treats coincidence counts, the draw volume calculator prepares the dose, the Poisson distribution calculator gives counting statistics and the dead time calculator corrects for count losses.