Nuclear medicine image counts and acquisition time calculator

Enter the target image noise, the number of pixels of the region, the camera sensitivity and the activity in the region to get the counts per pixel, the total counts, the count rate and the acquisition time, or the noise reached in a given time.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
%

Value out of range

Value out of range

MBq

Value out of range

cps/MBq

Value out of range

s

Value out of range

Results

Acquisition time for the target noise

229 s

A reasonable acquisition time
Counts per pixel needed
400
Total counts needed in the region
800,000
Count rate from the region
3,500 cps
Noise per pixel reached in the chosen time
4.36 %
Acquisition time in minutes
3.8 min

The nuclear medicine image counts calculator gives the counts per pixel, the total counts, the count rate and the acquisition time needed for a target image noise, or the noise reached in a chosen time, from the activity and the system sensitivity.

How the calculation works

  • Enter the target noise and the pixels of the region.
  • Enter the activity and the sensitivity.
  • Read the time and the noise in a chosen time.

Example

A noise of 5 % per pixel needs (100 / 5)² = 400 counts per pixel, so 800,000 counts over 2000 pixels. With 50 MBq in the region and 70 cps/MBq, the count rate is 3,500 cps, and the acquisition takes 229 s, or 3.8 min. In 300 s the noise would be 4.36 %.

Why it matters

The calculator helps nuclear medicine technologists, physicists and students who plan acquisitions and want to understand the link between activity, time and image noise.

Good to know

The model uses Poisson noise only, with no correction for background, scatter, attenuation or the image processing, which includes smoothing that reduces the visible noise at the expense of resolution. The sensitivity depends on the collimator, the photon energy, the depth of the source and the energy window. Only the counts in the region count, so a bigger region needs more counts for the same noise per pixel but the same noise for the total counts. The motion of the patient increases with the scan time, and decay during a long acquisition reduces the count rate, especially for short-lived radionuclides, so allow a margin. The time is only a planning estimate and does not replace the local protocol.

The PET count rate calculator treats coincidence counts, the draw volume calculator prepares the dose, the Poisson distribution calculator gives counting statistics and the dead time calculator corrects for count losses.

Frequently asked questions

Why does the noise follow the counts?

Radioactive decay is a random process that follows Poisson statistics, so the standard deviation of N counts is √N and the relative noise is 1 / √N. To halve the noise you need four times the counts.

What is the system sensitivity?

It is the count rate recorded per unit of activity, in counts per second per megabecquerel. A typical gamma camera with a general purpose collimator records about 70 to 100 cps/MBq for technetium-99m, which is about 150 to 200 counts per minute per microcurie.

How do I use the result?

Compare the time with the clinical protocol. Many static studies target 100 000 to 500 000 counts for the whole image, which gives only a few hundred counts per pixel in a typical matrix, so the noise per pixel is relatively high.