Detector dead time correction calculator

Enter the observed count rate and the dead time of the detector to get the true count rate, the percentage of counts lost and the maximum count rate, or enter the rates of two sources alone and together to estimate the dead time.

Updated on
Your inputs
counts/s

Value out of range

µs

Value out of range

counts/s

Value out of range

counts/s

Value out of range

counts/s

Value out of range

Results

True count rate (non-paralyzable model)

1,111 counts/s

A small correction, reliable for most work
Counts lost
10.0 %
Maximum observable count rate (1 / dead time)
10,000 counts/s
Dead time from the two-source method (approximation)
25.0 µs

The dead time correction calculator gives the true count rate of a detector from the observed rate and the dead time, the percentage of counts lost, and estimates the dead time with the two-source method.

How the calculation works

  • Enter the observed count rate and the dead time.
  • Optionally enter the rates of two sources alone and together.
  • Read the true rate, the loss and the estimated dead time.

Example

A Geiger counter with a dead time of 100 µs records 1000 counts per second. The true rate is 1000 / (1 − 1000 × 0.0001) = 1,111 counts per second, so 10.0 % of the events are lost. The maximum observable rate is 10,000 counts per second. For the two-source test, rates of 3000, 4000 and 6400 give a dead time of 25.0 µs.

Why it matters

The calculator helps health physicists, laboratory students, nuclear medicine staff and technicians who use Geiger counters, scintillation counters or well counters at high count rates.

Good to know

The non-paralyzable model assumes that events arriving during the dead time are simply lost and do not extend it. Geiger-Müller counters behave closer to the paralyzable model at very high rates, where the observed rate falls as the true rate rises, so a measured rate does not tell which side of the maximum you are on. Corrections above about 20 percent are not reliable, and it is better to move the source farther away or to reduce its activity. The two-source method also needs the same background in all three measurements and a well-defined geometry. Dead time is different from the recovery time and from the resolving time of a coincidence circuit. Counting statistics add an uncertainty to any corrected rate.

The Poisson distribution calculator gives counting statistics, the radioactive decay calculator treats activity, the minimum detectable activity calculator treats detection limits and the image counts calculator links counts and noise.

Frequently asked questions

What is dead time?

After each event a detector needs a short time to recover, during which it cannot record another event. At high rates events are lost. For a non-paralyzable detector the true rate is n = m / (1 − m τ), where m is the measured rate and τ the dead time.

What is the two-source method?

Two sources are counted separately and together. Because of the lost counts, the combined rate is lower than the sum, and the deficit gives the dead time: τ ≈ (R1 + R2 − R12) / (2 R1 R2). The method is approximate and needs a stable geometry and a small background.

Which dead times are typical?

A Geiger-Müller tube has about 100 to 300 µs, a scintillation counter about 1 µs, a proportional counter a few microseconds and a modern gamma camera several microseconds per event. Always use the value for your instrument.

Was this calculator helpful?