PET noise equivalent count rate (NECR) calculator

Enter the true, scattered and random coincidence rates and the randoms correction method to get the noise equivalent count rate, the scatter fraction, the random fraction and the number of counts needed for a target number of noise equivalent counts.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
kcps

Value out of range

kcps

Value out of range

kcps

Value out of range

million

Value out of range

Results

Noise equivalent count rate

50.0 kcps

A good share of the trues contributes effectively
Scatter fraction
28.6 %
Randoms fraction of the total coincidences
17.6 %
NECR as a share of the true rate
50 %
Time needed for the target noise equivalent counts
2,000 s

The PET noise equivalent count rate calculator gives the NECR, the scatter fraction and the randoms fraction from the rates of true, scattered and random coincidences, and the time needed for a target number of noise equivalent counts.

How the calculation works

  • Enter the trues, scatter and randoms rates.
  • Choose the randoms correction method.
  • Read the NECR, the fractions and the time.

Example

A scanner records 100 kcps of trues, 40 kcps of scatter and 30 kcps of randoms with a delayed window (k = 2). The NECR is 100² / (100 + 40 + 2 × 30) = 50.0 kcps, which is 50 % of the true rate. The scatter fraction is 28.6 % and the randoms fraction 17.6 %. To collect 100 million noise equivalent counts the scan lasts 2,000 s.

Why it matters

The calculator helps nuclear medicine physicists, PET technologists and students who compare scanners and choose the injected activity and the scan time.

Good to know

The NECR depends on the object, the scanner and the energy window, and its measurement follows standards such as the NEMA NU 2 protocol with a cylindrical phantom. Modern time-of-flight scanners gain an effective sensitivity beyond what the NECR shows, because the timing information reduces the noise propagation. A scatter fraction around 30 to 40 percent in 3D mode is usual, and randoms increase with the square of the activity in the field of view. A lower activity concentration at the time of scanning, for example after a long uptake time, reduces the randoms but also the signal. Use this page to understand the trade-off, not as a replacement for the performance measurements of a specific scanner.

The PET SUV calculator treats quantification, the image counts calculator links counts and noise, the contrast to noise calculator describes visibility and the dead time calculator treats count losses.

Frequently asked questions

What is the NECR?

The noise equivalent count rate is the rate of true coincidences that would give the same signal to noise ratio in an ideal system with no scatter and no randoms: NECR = T² / (T + S + kR). It is the standard figure of merit of the count rate performance of a PET scanner.

Why is k equal to 1 or 2?

If the randoms are estimated with a delayed coincidence window, the subtraction adds noise, and k = 2. If the randoms are estimated from the singles rates without extra noise, k = 1. The value used depends on the correction method of the scanner.

How does the NECR change with the activity?

At low activity the trues and scatter rise linearly and the randoms with the square of the activity, so the NECR increases, reaches a peak and then falls. The activity at the peak is the best compromise for the injected dose and the time after injection.