The radioiodine patient release calculator gives the dose rate at a distance, the dose to a person nearby and the waiting time before a patient treated with iodine-131 can leave the hospital, from the retained activity, the effective half-life and the occupancy.
How the calculation works
- Enter the retained activity and the effective half-life.
- Enter the distance, the occupancy factor and your limits.
- Read the dose rate, the dose and the waiting time.
Example
A patient retaining 555 MBq has a dose rate of 0.052 × 555 / 1² = 28.9 µSv/h at 1 m. With an effective half-life of 3 days and an occupancy of 0.25, a person nearby gets 28.86 × 72 / 0.693 × 0.25 = 0.75 mSv in total. The dose rate at 1 m falls below 20 µSv/h after 1.6 days.
Why it matters
The calculator helps nuclear medicine physicians, physicists, nurses and radiation protection officers who prepare release instructions, and students who learn how the dose to the public is estimated.
Good to know
This is a simplified model. It treats the patient as a point source in air, ignores the shielding by the body, which lowers the measured dose rate, and uses one effective half-life, whereas the retention often follows two components. The occupancy factor depends on the situation, and instructions to a patient include sleeping apart, avoiding close contact with children and pregnant women, hygiene measures and travel limits for several days. The release criteria are set by national rules, and in some cases the patient is hospitalized in a shielded room until the activity falls below the limit. A measured dose rate with a calibrated instrument at the time of release is the best basis for the decision. This page is educational only.
Related tools
The radioiodine activity calculator plans the treatment, the waste decay calculator treats storage, the inverse square law calculator shows the distance effect and the effective half-life calculator combines physical and biological clearance.