Size-specific dose estimate (SSDE) calculator

Enter the CTDIvol shown by the scanner, the anteroposterior and lateral dimensions of the patient and the phantom used (16 cm or 32 cm) to get the effective diameter, the AAPM Report 204 conversion factor and the size-specific dose estimate.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
mGy

Value out of range

cm

Value out of range

cm

Value out of range

Results

Size-specific dose estimate

14.42 mGy

The dose is higher than the displayed CTDIvol
Effective diameter
25.7 cm
Conversion factor (AAPM Report 204)
1.442
Change against the displayed CTDIvol
44 %

The size-specific dose estimate calculator converts the CTDIvol of a CT scan into a size-specific dose estimate using the effective diameter of the patient and the conversion factors of AAPM Report 204.

How the calculation works

  • Enter the CTDIvol from the scanner console.
  • Enter the anteroposterior and lateral dimensions and choose the phantom.
  • Read the effective diameter, the factor and the SSDE.

Example

A patient measuring 22 cm front to back and 30 cm wide has an effective diameter of √(22 × 30) = 25.7 cm. For the 32 cm phantom the factor is 3.7044 × exp(−0.03672 × 25.7) = 1.442. With a CTDIvol of 10 mGy the SSDE is 14.42 mGy, about 44 % above the displayed value.

Why it matters

The calculator helps CT technologists, medical physicists and radiologists who track the dose of children and adults of different sizes, and students who learn why a dose index is not a patient dose.

Good to know

The factors come from fits to the tables of AAPM Report 204 and are valid for effective diameters from about 8 to 38 cm. The report gives separate fits for the 16 cm and 32 cm reference phantoms, and the choice of phantom depends on the scanner and the protocol, so check the console or the manual. For head scans the 16 cm phantom is normally the reference, and many pediatric body protocols use the 16 cm phantom as well. The SSDE does not account for the attenuation of the actual tissues, such as the lungs, which differ from the water-like phantom, and a correction using water equivalent diameter is also described in later reports. Use SSDE to compare doses of patients of different sizes and to set size-based protocols.

The CT dose calculator gives the CTDIvol and the DLP, the Hounsfield unit calculator links attenuation and CT numbers, the dose comparison calculator puts a dose in context and the contrast to noise calculator relates dose to image quality.

Frequently asked questions

What is the SSDE?

The size-specific dose estimate corrects the CTDIvol, which is measured in a standard phantom, for the size of the patient. A small patient absorbs a higher dose than the phantom for the same scanner output, a large patient a lower one.

How is the effective diameter found?

It is the diameter of a circle with the same area as the cross-section of the patient, approximately the square root of the product of the anteroposterior and lateral dimensions, measured on the images or the localizer.

Is the SSDE the dose to the patient?

It is an estimate of the average dose in the scanned volume, not of an organ dose and not of the effective dose. It does not replace dosimetry for special cases, such as pregnancy or very high dose procedures.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.