Standard addition calculator

Enter the signal of the sample, the signal after adding a known amount of standard, and the volumes and concentration of the standard to get the concentration in the sample, corrected for dilution.

Updated on
Your inputs

Value out of range

Value out of range

mg/L

Value out of range

mL

Value out of range

mL

Value out of range

Results

Concentration in the sample

5.8962 mg/L

Concentration added by the spike
4.5455 mg/L
Spike as a share of the sample level
77.1 %
Spike of a suitable size
Sample signal corrected for dilution
0.2273
Sensitivity of the method
0.04240 signal per mg/L
Signal gain from the spike
1.680 ×

The standard addition calculator gives the concentration of an analyte in a sample from one spike of standard, with dilution correction, the sensitivity and a check on the size of the spike.

How the calculation works

  • Enter the signal of the sample and the signal after the spike.
  • Enter the concentration and the volume of the standard, and the volume of the sample.
  • The calculator gives the concentration, the added concentration, the spike share, the corrected signal and the sensitivity.

Example

A sample gives a signal of 0.250. After adding 1 mL of a 50 mg/L standard to 10 mL of sample, the signal is 0.420. The denominator is 0.420 × 11 − 0.250 × 10 = 2.12 and the numerator 0.250 × 50 × 1 = 12.5, so the sample holds 5.8962 mg/L. The spike added 4.5455 mg/L, which is 77.1 % of the sample level.

Why it matters

The calculator helps analytical chemistry students, laboratory technicians who measure metals by atomic absorption or ICP, and environmental or food analysts who work with complex matrices.

Good to know

The method corrects for multiplicative matrix effects, where the sample changes the sensitivity, but not for additive interferences such as a background signal. Subtract the blank first. It assumes a linear response between the two points and a signal that goes through zero for zero analyte. Mix well and let the spiked sample stabilize before measuring. When accuracy matters, use three to five additions and fit a line, which also gives an estimate of the uncertainty. The method is slower than an external calibration, since each sample needs its own spikes, so keep it for difficult matrices such as wine, soil extracts, blood, seawater or industrial effluents.

The calibration curve calculator uses external standards, the limit of detection calculator gives the lowest measurable level, the dilution calculator handles volumes and concentrations and the Beer-Lambert calculator links absorbance and concentration.

Frequently asked questions

What is the standard addition method?

You measure the sample, add a known amount of the analyte to it and measure again. The increase in signal gives the response per unit of concentration in the sample itself, so matrix effects that change the sensitivity cancel out. It is used when a normal calibration curve would be biased.

How is the concentration computed?

With a single spike, Cx = S0 × Cs × Vs / (S1 × (Vx + Vs) − S0 × Vx). The spiked signal is measured in a larger volume, so the formula corrects for dilution. With several spikes, plot the signal against the added concentration and read the intercept on the concentration axis.

How large should the spike be?

Aim for an addition that raises the signal by roughly 50 % to 200 % of the original reading, and keep the volume small, below about 10 % of the sample, to limit dilution. Stay within the linear range of the instrument.