Mayneord F factor calculator

Enter the percentage depth dose known at a reference distance, the depth, the depth of maximum dose and the two distances to get the Mayneord F factor and the corrected percentage depth dose at the new distance.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
%

Value out of range

cm

Value out of range

cm

Value out of range

cm

Value out of range

cm

Value out of range

Results

Mayneord F factor

0.9631

Corrected PDD at the new distance
64.34 %
Change in dose at depth
-3.69 %
Small to moderate change: the approximation is reasonable
Inverse-square ratio at the reference distance
0.8514
Inverse-square ratio at the new distance
0.8200
Difference in distance
-20 cm

The Mayneord F factor calculator corrects a percentage depth dose for a change in source-to-surface distance, from the depth, the depth of maximum dose and the two distances.

How the calculation works

  • Enter the PDD at the reference distance, the depth and the depth of maximum dose.
  • Enter the reference and new distances.
  • The calculator gives the F factor, the corrected PDD and the change in dose.

Example

A PDD of 66.8 % at 10 cm depth for a distance of 100 cm needs to be converted to 80 cm, with dmax at 1.5 cm. F = [(80 + 1.5) / (80 + 10)]² / [(100 + 1.5) / (100 + 10)]² = 0.8200 / 0.8514 = 0.9631. The corrected PDD is 66.8 × 0.9631 = 64.34 %, a change of -3.69 %.

Why it matters

The calculator helps radiation therapy students, dosimetrists and medical physicists who need to adapt tabulated data to a different treatment distance.

Good to know

The method assumes that the scatter contribution to the PDD does not change with the distance, which is not exact for large field sizes and low energies. Some centers use a more detailed version that includes the field size at the surface. Official data should come from measured tables or the treatment planning system. The F factor is not a change in the total dose to the patient: monitor units still need correction for the inverse-square change at dmax. Check units, since distances are measured from the source to the surface, not to the isocenter, in this technique. Treat the result as a first estimate and confirm it with measured depth dose tables for the actual beam.

The monitor unit calculator turns a prescription into machine output, the inverse square law calculator shows the distance effect, the BED calculator compares schedules and the brachytherapy calculator treats seed implants.

Frequently asked questions

What is the Mayneord F factor?

It is a correction that converts a percentage depth dose measured at one source-to-surface distance to another by accounting for the change in the inverse-square dose gradient, assuming that scatter does not change: F = [(S2 + dmax) / (S2 + d)]² / [(S1 + dmax) / (S1 + d)]².

When is it accurate?

For modest changes of distance, a few centimeters to about 10 cm. For large changes, or for low-energy beams where scatter is important, it loses accuracy, and measured data should be used.

Does the dose go up or down?

A longer distance gives a higher percentage depth dose, because the inverse-square gradient is gentler. A shorter distance gives a lower one. Typical effect: about 1 % to 2 % per 10 cm for deep points.