The MRI scan time calculator gives the acquisition time of a sequence from the TR, the phase encoding steps, the number of averages, the turbo factor, the parallel imaging factor and the 3D partitions, and the time saved by each technique.
How the calculation works
- Enter the TR, the phase encoding steps and the NEX.
- Enter the turbo factor, the parallel imaging factor and the partitions.
- Read the acquisition time and the time saved.
Example
A sequence with a TR of 3000 ms, 256 phase steps and 2 averages would take 3000 × 256 × 2 = 1536 s (25.6 min) in conventional spin echo. With a turbo factor of 8 and a parallel imaging factor of 2 the time is 1536 / 16 = 96 s, or 1.60 minutes, saving 24.0 minutes.
Why it matters
The calculator helps MRI technologists, radiology residents, medical physicists and students who plan protocols and want to understand why a scan lasts as long as it does.
Good to know
Real scanners add overheads such as preparation pulses, dummy cycles, navigators or breath-hold gaps, and the total protocol time includes the localizer, the coil placement and the contrast injection. For 2D multi-slice imaging the number of slices that fit in one TR is limited by the time per slice, and extra slices require more acquisitions. Partial Fourier and reduced matrix techniques also shorten the scan by acquiring only a part of k-space. The square-root rule for the signal to noise ratio shown here ignores the specific loss of parallel imaging, so the real signal to noise ratio is lower. Choose the combination of acceleration and averages that keeps the image quality needed, and consider patient motion: a shorter scan is often sharper in practice.