MRI scan time calculator

Enter the repetition time, the number of phase encoding steps, the number of averages, the echo train length, the parallel imaging factor and the number of 3D partitions to get the acquisition time and the gain from each acceleration technique.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
ms

Value out of range

Value out of range

Value out of range

Value out of range

Value out of range

Value out of range

Results

Acquisition time

96 s

A short acquisition, comfortable for the patient
Acquisition time in minutes
1.60 min
Time without turbo factor or parallel imaging
25.6 min
Time saved by the acceleration
24.0 min
Relative SNR against the unaccelerated scan (square-root law only)
25 %

The MRI scan time calculator gives the acquisition time of a sequence from the TR, the phase encoding steps, the number of averages, the turbo factor, the parallel imaging factor and the 3D partitions, and the time saved by each technique.

How the calculation works

  • Enter the TR, the phase encoding steps and the NEX.
  • Enter the turbo factor, the parallel imaging factor and the partitions.
  • Read the acquisition time and the time saved.

Example

A sequence with a TR of 3000 ms, 256 phase steps and 2 averages would take 3000 × 256 × 2 = 1536 s (25.6 min) in conventional spin echo. With a turbo factor of 8 and a parallel imaging factor of 2 the time is 1536 / 16 = 96 s, or 1.60 minutes, saving 24.0 minutes.

Why it matters

The calculator helps MRI technologists, radiology residents, medical physicists and students who plan protocols and want to understand why a scan lasts as long as it does.

Good to know

Real scanners add overheads such as preparation pulses, dummy cycles, navigators or breath-hold gaps, and the total protocol time includes the localizer, the coil placement and the contrast injection. For 2D multi-slice imaging the number of slices that fit in one TR is limited by the time per slice, and extra slices require more acquisitions. Partial Fourier and reduced matrix techniques also shorten the scan by acquiring only a part of k-space. The square-root rule for the signal to noise ratio shown here ignores the specific loss of parallel imaging, so the real signal to noise ratio is lower. Choose the combination of acceleration and averages that keeps the image quality needed, and consider patient motion: a shorter scan is often sharper in practice.

The spin echo calculator treats contrast, the Ernst angle calculator treats gradient echo, the chemical shift calculator treats artifacts and the contrast to noise calculator treats visibility.

Frequently asked questions

What is the scan time formula?

For a conventional 2D sequence, TA = TR × number of phase encoding steps × NEX. Each excitation fills one line of k-space; with a 3D sequence the time is multiplied by the number of partitions. All slices of a 2D multi-slice acquisition share the same TR when it is long enough.

How do turbo and parallel imaging save time?

A turbo (fast) spin echo acquires several lines of k-space per TR, divided by the echo train length. Parallel imaging uses the sensitivity of several coil elements to skip lines, so the time is divided by the acceleration factor.

What do we pay for a shorter scan?

The signal to noise ratio falls with the square root of the acquisition time, plus an extra loss with parallel imaging due to the geometry of the coil. Long echo trains also blur the image and change the contrast.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.