The MRI spin echo signal calculator gives the relative signal of two tissues and their contrast for a chosen TR and TE, from the proton density, T1 and T2 of each, and the echo time that maximizes the T2 contrast.
How the calculation works
- Enter the TR and the TE.
- Enter the proton density, T1 and T2 of the two tissues.
- Read the signals, the contrast and the best TE.
Example
With a TR of 500 ms and a TE of 15 ms, gray matter (PD 80, T1 900, T2 100) gives a signal of 29.35 and white matter (PD 70, T1 600, T2 80) gives 32.81. The contrast is 11.1 % and the white matter appears brighter, as on a T1-weighted brain image. The echo time for the best T2 contrast between them is 89.3 ms.
Why it matters
The calculator helps MRI technologists, radiology residents, medical physicists and students who learn how the sequence parameters change the image contrast.
Good to know
The formula describes an ideal spin echo with a 90 degree excitation and a 180 degree refocusing pulse. Real images also depend on the inhomogeneity of the coils, the slice profile and the flip angles, and turbo spin echo sequences add a dependence on the echo train length. The T1 and T2 values quoted are approximate and vary with the field strength, the temperature and the measurement method. The contrast shown is a ratio of signals and ignores noise, so a high contrast with a low signal may still be hard to see. A real protocol is a compromise between contrast, signal to noise and scan time, and the optimal TE for the T2 contrast is only one choice. Use the page to understand the logic of weighting, not to design a protocol.
Related tools
The Ernst angle calculator treats gradient echo sequences, the scan time calculator links parameters and duration, the chemical shift calculator treats fat-water artifacts and the contrast to noise calculator describes visibility.