MRI spin echo signal and contrast calculator

Enter the repetition time and the echo time, then the proton density, T1 and T2 of two tissues to get the spin echo signal of each, their contrast and the echo time that maximizes the T2 contrast.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
ms

Value out of range

ms

Value out of range

%

Value out of range

ms

Value out of range

ms

Value out of range

%

Value out of range

ms

Value out of range

ms

Value out of range

Results

Relative signal of tissue A

29.35

Relative signal of tissue B
32.81
Contrast between the tissues
11.1 %
A T1-weighted setting: short TR and short TE
Echo time that maximizes the T2 contrast
89.3 ms
T1 recovery of tissue A (share of full magnetization)
43 %

The MRI spin echo signal calculator gives the relative signal of two tissues and their contrast for a chosen TR and TE, from the proton density, T1 and T2 of each, and the echo time that maximizes the T2 contrast.

How the calculation works

  • Enter the TR and the TE.
  • Enter the proton density, T1 and T2 of the two tissues.
  • Read the signals, the contrast and the best TE.

Example

With a TR of 500 ms and a TE of 15 ms, gray matter (PD 80, T1 900, T2 100) gives a signal of 29.35 and white matter (PD 70, T1 600, T2 80) gives 32.81. The contrast is 11.1 % and the white matter appears brighter, as on a T1-weighted brain image. The echo time for the best T2 contrast between them is 89.3 ms.

Why it matters

The calculator helps MRI technologists, radiology residents, medical physicists and students who learn how the sequence parameters change the image contrast.

Good to know

The formula describes an ideal spin echo with a 90 degree excitation and a 180 degree refocusing pulse. Real images also depend on the inhomogeneity of the coils, the slice profile and the flip angles, and turbo spin echo sequences add a dependence on the echo train length. The T1 and T2 values quoted are approximate and vary with the field strength, the temperature and the measurement method. The contrast shown is a ratio of signals and ignores noise, so a high contrast with a low signal may still be hard to see. A real protocol is a compromise between contrast, signal to noise and scan time, and the optimal TE for the T2 contrast is only one choice. Use the page to understand the logic of weighting, not to design a protocol.

The Ernst angle calculator treats gradient echo sequences, the scan time calculator links parameters and duration, the chemical shift calculator treats fat-water artifacts and the contrast to noise calculator describes visibility.

Frequently asked questions

What is the spin echo signal equation?

For a spin echo, the signal is S = PD × (1 − exp(−TR / T1)) × exp(−TE / T2). The first factor is the recovery of the longitudinal magnetization between pulses, the second is the decay of the transverse magnetization before the echo.

How do TR and TE set the weighting?

A short TR and a short TE emphasize differences in T1 (fat bright, fluid dark). A long TR and a long TE emphasize T2 (fluid bright). A long TR with a short TE shows proton density. Mid values mix the effects.

Where do the T1 and T2 values come from?

They depend on the tissue and on the field strength. At 1.5 T, white matter has a T1 near 600 ms and a T2 near 80 ms, gray matter 900 and 100 ms, and cerebrospinal fluid more than 3000 and 2000 ms. T1 grows with the field strength.

Was this calculator helpful?

Thanks for your feedback!

Results are estimates provided for information purposes. They do not replace professional advice.