MRI chemical shift artifact calculator

Enter the field strength, the receiver bandwidth, the number of frequency encoding points and the field of view to get the fat-water frequency difference, the displacement in pixels and millimeters and the bandwidth needed to keep the shift within one pixel.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
T

Value out of range

ppm

Value out of range

kHz

Value out of range

Value out of range

mm

Value out of range

Results

Fat-water shift in pixels

3.58

A large displacement: increase the bandwidth per pixel
Fat-water frequency difference
447 Hz
Bandwidth per pixel
125 Hz
Displacement in millimeters
3.49 mm
Total bandwidth needed for a shift of one pixel
114.4 kHz
Relative SNR at the bandwidth needed (against the present bandwidth)
53 %

The MRI chemical shift artifact calculator gives the fat-water frequency difference, the displacement in pixels and in millimeters, and the bandwidth needed to keep the shift within one pixel, from the field strength, the receiver bandwidth, the matrix and the field of view.

How the calculation works

  • Enter the field strength and the chemical shift.
  • Enter the receiver bandwidth, the frequency points and the field of view.
  • Read the shift in pixels, the displacement and the bandwidth needed.

Example

At 3 T the proton frequency is 127.7 MHz, so 3.5 ppm is 447 Hz. With a bandwidth of 32 kHz over 256 points the bandwidth per pixel is 125 Hz, so the shift is 3.58 pixels, or 3.49 mm for a 250 mm field of view. To keep it within one pixel the total bandwidth must be 114.4 kHz, at the price of a signal to noise ratio of 53 %.

Why it matters

The calculator helps MRI technologists, radiology residents, medical physicists and students who understand artifacts, and anyone who sets up sequences in the abdomen, the spine or the orbit where fat and water meet.

Good to know

The artifact appears only in the frequency encoding direction, so swapping the phase and frequency directions can move it away from the region of interest. It is larger at high field, with a low bandwidth and with a large matrix. Echo planar imaging has a much larger shift in the phase direction because the bandwidth per pixel is very low, which is why fat suppression is almost always used. Chemical shift imaging can also be used on purpose to detect fat in an adrenal lesion through the signal loss on opposed-phase images. The proton gyromagnetic ratio used is 42.577 MHz per tesla. Always check the bandwidth per pixel in the parameters of the protocol, since some vendors display the total bandwidth and others the bandwidth per pixel.

The scan time calculator links parameters and duration, the spin echo calculator treats contrast, the Ernst angle calculator treats gradient echo and the Nyquist sampling calculator treats sampling and aliasing.

Frequently asked questions

What is the chemical shift artifact?

Protons in fat and in water resonate at slightly different frequencies, about 3.5 ppm apart. Since the scanner locates the signal by its frequency, fat appears shifted along the frequency encoding direction compared with water, which gives bright and dark bands at their boundaries.

Why does it grow with the field strength?

The frequency difference in hertz is proportional to the field strength: 3.5 ppm of 63.9 MHz at 1.5 T is about 224 Hz, and about 447 Hz at 3 T. For the same bandwidth per pixel, the shift in pixels doubles when the field doubles.

What is the cost of a larger bandwidth?

The noise rises with the square root of the receiver bandwidth, so the signal to noise ratio falls. A compromise is made between the artifact and the noise, and fat suppression removes the problem for most applications.