Ernst angle calculator (gradient echo MRI)

Enter the repetition time, the T1 and T2* of a tissue, the echo time and the flip angle of a gradient echo sequence to get the Ernst angle, the signal at your flip angle, the maximum signal and the share of the maximum reached.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
ms

Value out of range

ms

Value out of range

ms

Value out of range

ms

Value out of range

°

Value out of range

Results

Ernst angle

14.7 °

Relative signal at your flip angle
0.0786
Maximum signal at the Ernst angle
0.1005
Share of the maximum signal reached
78 %
A noticeable loss against the optimal signal
T2* decay factor at the echo time
0.779

The Ernst angle calculator gives the flip angle that maximizes the signal of a spoiled gradient echo sequence from the TR and the T1, and the signal at the flip angle you chose, with the T2* decay at the echo time.

How the calculation works

  • Enter the TR, the T1, the TE and the T2*.
  • Enter your flip angle.
  • Read the Ernst angle, the signal and the share of the maximum.

Example

For a TR of 30 ms and a T1 of 900 ms, exp(−30 / 900) = 0.9672, so the Ernst angle is acos(0.9672) = 14.7°. At a flip angle of 30°, with a TE of 10 ms and a T2* of 40 ms, the relative signal is 0.0786, which is 78 % of the maximum signal of 0.1005.

Why it matters

The calculator helps MRI technologists, radiology residents, medical physicists and students who learn gradient echo imaging, and researchers who set up fast 3D or functional sequences.

Good to know

The equation applies to a spoiled gradient echo sequence in the steady state with ideal spoiling, and it neglects the effect of the slice profile and of B1 inhomogeneity, which makes the actual flip angle vary across the image. Balanced steady-state sequences follow a different formula that also depends on T2. The T1 depends on the field strength and the tissue, and the Ernst angle is therefore different for gray matter, white matter, fat or a contrast-enhanced lesion. The signal shown is relative to the equilibrium magnetization and is not corrected for coil sensitivity or receiver gain. Because the signal near the Ernst angle varies slowly, a small error in the angle costs little signal, which makes the choice forgiving.

The spin echo calculator treats the T1 and T2 weighting, the scan time calculator links parameters and duration, the chemical shift calculator treats fat-water artifacts and the signal to noise calculator treats noise.

Frequently asked questions

What is the Ernst angle?

It is the flip angle that gives the maximum signal in a spoiled gradient echo sequence for a given TR and T1: cos α = exp(−TR / T1). A short TR compared with T1 needs a small flip angle, a long TR needs an angle close to 90°.

Why use a small flip angle?

With a short TR, a 90° pulse would saturate the signal because the longitudinal magnetization cannot recover. A smaller angle keeps part of the magnetization, so the steady-state signal is larger and the scan can be much faster.

Does the Ernst angle give the best contrast?

Not necessarily. It maximizes the signal of one tissue, not the contrast between two. Larger angles give more T1 weighting, smaller angles more proton density weighting. In practice the angle is chosen for the contrast wanted.