Ultrasound resolution calculator

Enter the frequency, the number of cycles per pulse, the speed of sound, the aperture and the focal length of a transducer to get the wavelength, the axial resolution, the lateral resolution at the focus, the near field length and the divergence of the far field.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MHz

Value out of range

Value out of range

m/s

Value out of range

mm

Value out of range

mm

Value out of range

Results

Axial resolution

0.308 mm

Lateral resolution at the focus
1.232 mm
Wavelength in tissue
0.308 mm
Near field length of an unfocused disc
182.6 mm
The natural near field is longer than the focal length
Far field divergence half-angle
1.44 °
Lateral to axial resolution ratio
4.0

The ultrasound resolution calculator gives the wavelength, the axial resolution, the lateral resolution at the focus, the near field length and the far field divergence of a transducer from its frequency, cycles per pulse, aperture and focal length.

How the calculation works

  • Enter the frequency, the cycles per pulse and the speed of sound.
  • Enter the aperture and the focal length.
  • Read the axial and lateral resolution, and the near field.

Example

At 5 MHz in tissue at 1540 m/s the wavelength is 0.308 mm. A pulse of 2 cycles gives an axial resolution of 2 × 0.308 / 2 = 0.308 mm. With an aperture of 15 mm focused at 60 mm, the lateral resolution is 0.308 × 60 / 15 = 1.232 mm, about 4.0 times worse than the axial one. The natural near field is 182.6 mm long.

Why it matters

The calculator helps sonography students, technologists, biomedical engineers and medical physicists who choose a probe or study image quality, and teachers who explain why ultrasound images are sharper along the beam than across it.

Good to know

The formulas are first-order estimates for a simple transducer. Real beams have side lobes and grating lobes, tissue changes the beam, and the lateral resolution varies with depth: it is best at the focus and wider elsewhere. Electronic focusing in array transducers adjusts the focus during transmission and reception, and dynamic receive focusing extends the zone of good resolution. The elevation (slice thickness) resolution, perpendicular to the image plane, is often the worst and is set by the lens of the probe. Broadband pulses with few cycles give better axial resolution but spread the energy over a wider band, which reduces the sensitivity. Harmonic imaging improves the lateral resolution because the harmonic beam is narrower. Always check the specifications of the probe.

The pulse-echo timing calculator links depth and frame rate, the mechanical index calculator checks the output, the acoustic impedance calculator treats reflection and the MTF calculator describes sharpness in general.

Frequently asked questions

What is axial resolution?

It is the smallest distance along the beam between two reflectors that can be seen separately. It equals half the spatial pulse length, which is the number of cycles times the wavelength, divided by two. Higher frequencies and shorter pulses give a better axial resolution.

What sets the lateral resolution?

It is about the width of the beam at the focus, roughly the wavelength times the focal length divided by the aperture. A wider aperture, a shorter focal length and a higher frequency narrow the beam. Beyond the focus the beam diverges and the resolution worsens.

Why not always use the highest frequency?

Attenuation grows with frequency, about 0.5 dB/cm/MHz in soft tissue, so high frequencies lose penetration. The choice is a compromise: 2 to 5 MHz for deep abdominal and cardiac imaging, 7 to 15 MHz for superficial structures and above for skin and eye.

Was this calculator helpful?