The chromatographic resolution calculator gives the resolution of two peaks, the selectivity, the retention factors and the number of plates needed for a baseline separation.
How the calculation works
- Enter the retention times and the base widths of the two peaks.
- Enter the dead time.
- The calculator gives the resolution, the selectivity, the retention factors and the plates needed for Rs = 1.5.
Example
Two peaks elute at 5.2 and 5.9 min with base widths of 0.25 and 0.27 min. The resolution is 2 × 0.7 / 0.52 = 2.69. With a dead time of 1 min, the retention factors are 4.20 and 4.90, and the selectivity is 1.167. A resolution of 1.5 would need about 2,557 plates for this selectivity, much less than the column already provides.
Why it matters
The calculator helps chemistry students, analysts and method developers who judge a separation in liquid or gas chromatography and decide whether to change the column or the mobile phase.
Good to know
A resolution above 1.5 is the usual goal for quantitative work, and 2.0 gives a margin for variation. The widths must be measured in the same way for the two peaks, for example at the base by tangents. When the two peaks are of very different sizes, a larger resolution is needed, because the large peak’s tail hides the small one. The plates needed assume the two retention factors given here. For the best use, also keep the retention factor between 2 and 10, since below 1 the compounds elute too near the void time and above 10 the analysis is slow. Gradient elution changes the picture and calls for other formulas.
Related tools
The plate number calculator gives the column efficiency, the Rf value calculator covers thin layer chromatography, the van Deemter calculator relates flow and plate height and the standard deviation calculator describes spread.