Electron beam depth rules calculator

Enter the nominal electron energy and the depth of the target to get the rule-of-thumb depths of the 90 %, 80 % and 50 % doses, the practical range, the lead shield thickness and the energy needed to cover the target.

Updated on

Educational tool, not medical advice: this page does not diagnose and does not replace a doctor, nurse or pharmacist. It has not been reviewed by a healthcare professional. Editorial and medical policy

Your inputs
MeV

Value out of range

cm

Value out of range

Results

Depth of the 90 % dose (therapeutic range)

3.75 cm

Depth of the 80 % dose
4.29 cm
Depth of the 50 % dose
5.15 cm
Practical range
6.00 cm
Energy needed to cover the target with 90 %
9.6 MeV
The chosen energy covers the target at the 90 % level
Lead thickness to stop the beam
6.0 mm
Margin between the energy chosen and the energy needed
2.4 MeV

The electron beam depth rules calculator gives the depth of the 90 %, 80 % and 50 % doses, the practical range, the lead shield thickness and the energy needed to cover a target from the nominal electron energy.

How the calculation works

  • Enter the nominal electron energy in MeV.
  • Enter the depth of the distal edge of the target.
  • The calculator gives the depths, the range, the lead thickness, the energy needed and the margin.

Example

A 12 MeV beam has a 90 % depth of 12 / 3.2 = 3.75 cm, an 80 % depth of 12 / 2.8 = 4.29 cm, a 50 % depth of 5.15 cm and a practical range of 12 / 2 = 6.00 cm. A target ending at 3 cm needs 3 × 3.2 = 9.6 MeV, so 12 MeV leaves a margin of 2.4 MeV. A lead shield of 6.0 mm stops the beam.

Why it matters

The calculator helps radiation therapy students and dosimetrists who learn the choice of electron energy for superficial tumors, skin cancers and boost fields.

Good to know

The rules apply to broad beams in water at the standard distance and are approximate. The actual depth doses depend on the applicator, the field size, the distance and tissue inhomogeneities such as bone, lung or air cavities. The surface dose is high, between 75 % and 95 %, and rises with energy. Bolus is used to bring the build-up closer to the skin and to flatten the surface. Always use measured data from the machine and the planning system. This page is for education only and is not a substitute for clinical dosimetry.

The Mayneord F factor calculator corrects depth dose for distance, the monitor unit calculator turns prescriptions into machine output, the BED calculator compares fractionation and the half-life calculator treats isotope decay.

Frequently asked questions

What are the depth rules for electron beams?

In water, the depth of the 90 % dose in centimeters is about the energy in MeV divided by 3.2, the 80 % depth about E / 2.8, the 50 % depth about E / 2.33 and the practical range about E / 2. They are rules of thumb valid between about 6 and 20 MeV.

How do I choose the energy?

Choose an energy whose 90 % depth reaches the distal edge of the target, since the dose drops quickly beyond it. The calculator multiplies the depth by 3.2 to give the minimum nominal energy, and the planner then picks the next available energy.

Why is the lead thickness E / 2?

As a rule, a lead thickness in millimeters equal to half of the energy in MeV stops the electrons and limits the bremsstrahlung to a small dose. Extra thickness is often added for safety, and in practice, the protection is designed by the physics team.