The radiotherapy vault primary barrier calculator gives the transmission factor and the thickness of a primary barrier from the design dose limit, the distance, the workload, the use and occupancy factors and the tenth-value layers of the material.
How the calculation works
- Enter the dose limit, the distance and the workload.
- Enter the use and occupancy factors.
- Enter the tenth-value layers and read the thickness.
Example
A weekly limit of 0.1 mSv at 5 m from a linear accelerator with a workload of 1000 Gy per week at 1 m, a use factor of 0.25 and full occupancy needs a transmission of 0.0001 × 25 / (1000 × 0.25 × 1) = 0.000010. This is 5.00 tenth-value layers. With 370 mm for the first and 330 mm for the next, the thickness of concrete is 1,690 mm, or 1.69 m.
Why it matters
The calculator helps medical physicists, radiation protection officers, architects and students who learn how a treatment room is shielded, and anyone who wants to see the effect of distance, use and occupancy on the wall thickness.
Good to know
The design goals and the factors are fixed by the regulator and by guidance such as NCRP Report 151 or national standards, and they differ by country. A typical design goal is 0.1 mSv per week in controlled areas and 0.02 mSv per week in public areas. The workload depends on the number of patients, the dose per patient and the treatment technique, and techniques such as IMRT raise the leakage workload. The distance is measured from the isocenter or the target to the point 0.3 m beyond the barrier. Tenth-value layers depend on the energy and the material, so use the values of your guidance. Always have the final design checked by a qualified expert and verified by a survey after construction.
Related tools
The half-value layer calculator treats attenuation by thickness, the time, distance and shielding calculator applies ALARA, the monitor unit calculator links machine output and dose and the inverse square law calculator shows the distance effect.